TANVI
Home / News / Filtration News / Analysis of the usage method of glove integrity tester
Analysis of the usage method of glove integrity tester
TANVI2024-09-12 TANVI327
The analysis of the usage method of the glove integrity tester mainly involves its operation process, precautions, and maintenance. The following is a detailed analysis of its usage method:


1、 Operation process


Preparation stage:


Before using the glove integrity tester, it is necessary to prepare the gloves and tester to be tested. Ensure that the gloves are not damaged or broken, and that the power supply, sensors, and other components of the detector are functioning properly.


Carefully read the operating manual of the detector and master the usage method proficiently.


Check whether the detector is intact, including whether there are damaged or loose parts. If there are any problems, contact after-sales maintenance in a timely manner.


Setting phase:


Power on and enter the main operation interface, select super administrator, administrator, or user login according to your needs.


Super administrators can check device parameter settings, set user permissions, and select appropriate testing modes (online or offline).


Enter the relevant parameters required for the test, such as inflation pressure value, etc.


Testing phase:


Put the gloves on the inflation device of the detector, ensuring that the gloves are sealed properly around the edges.


Activate the inflation device and inject a certain pressure of gas into the interior of the glove. The inflation pressure should be adjusted according to the material and purpose of the gloves, and it is generally recommended to be between 20-30kPa.


After inflation is complete, remove the gloves from the inflation device and place them on the detection device of the detector. Activate the detection device, and the sensor will start detecting gas leaks outside the gloves.


Determine the integrity of the gloves based on the display results of the detector. If gas leakage is found in the gloves, the detector will issue an alarm prompt.


Record and organize:


For each glove tested, its test results should be recorded, including information such as completeness and leakage location.


After the test is completed, you can choose to print the data, deflate it first, remove the sealing ring, take off the gloves, and cycle the test in sequence.


2、 Precautions


Safe operation:


During the testing process, it is necessary to ensure that gloves are fully worn and avoid placing unused gloves on the testing bench for testing.


The finger part of the glove should be placed in the cuff to avoid touching the electrode end to be tested.


Follow the operating procedures to ensure the normal operation of the equipment and the accuracy of the test results.


Environmental conditions:


The detector should be placed on a stable surface and ensure that the instrument is balanced and stable.


Pay attention to environmental conditions such as temperature, humidity, etc., and avoid using the equipment in harsh environments.


Maintenance and upkeep:


Regularly maintain and upkeep the glove integrity tester, including cleaning the equipment, checking for damage to components such as power and sensors.


Use a dry and clean cloth to wipe the surface of the instrument, avoiding the use of corrosive solvents or cleaning solutions.


When not in use for a long time, the instrument should be stored in a dry and dust-proof environment.


3、 Summary


The usage method of the glove integrity tester includes preparation stage, setup stage, testing stage, and recording and organizing steps. During use, attention should be paid to safe operation, environmental conditions, and equipment maintenance and upkeep. By using the glove integrity tester correctly, the integrity and safety of gloves can be ensured, thereby ensuring the safety and stability of experiments and production.